Task: monitor oil temperature, coolant, exhaust gas
Solution:
- Oil temperature: Pt100 with 4-wire (±0.2 °C accuracy for standard compliance)
- Coolant in/out: 2× Pt100 (calculate efficiency)
- Exhaust gas: robust type K thermocouple up to 800 °C
Temperature is the most frequently measured physical quantity in industry and research. This guide explains the key measurement methods, shows when to use which sensor, how data acquisition works correctly, and what often goes wrong in practice, with concrete application examples from development and production.
Whether in development, quality assurance, or production, temperature measurement serves many critical functions in industrial applications. Here are the main areas at a glance:
Practical tip: Temperature is often the first indicator of problems. A motor that suddenly runs 20 °C hotter than usual usually has a mechanical problem, long before other measured values stand out.
Temperatures can be captured either through direct thermal contact with the component or without contact via infrared. The choice depends on the process, accuracy, dynamics, and accessibility.
The first important decision: should the sensor touch the component or not?
Advantages: high accuracy, measures the actual component temperature
Disadvantages: the sensor has to be mounted and can affect the system
Examples: thermocouple in the engine block, Pt100 in the cooling circuit, NTC sensor on the circuit board
Advantages: no intervention in the system, very fast, suitable for moving or hazardous parts
Disadvantages: measures surface temperature only; the reading depends heavily on the surface finish (bare metal "radiates" differently than black paint)
Examples: pyrometer for rotating brake discs, thermal imaging camera for control cabinets
At a glance:
Contact measurement = high accuracy at the desired measuring point; IR measurement = fast, contactless, flexible, but mind the emissivity
The following sensor types cover most industrial applications. They differ in measuring principle, accuracy, dynamics, robustness, and integration effort.
Thermocouples are widespread and the "classic" for harsh environments, wide temperature ranges, and fast response. With their simple two-wire connection and standard connectors (thermocouple plugs), they are easy to replace and easy to source. They are also relatively inexpensive.
Thermocouples consist of two conductors made of different metals that are joined at the sensor tip. This is the measuring junction. Due to the Seebeck effect, a small voltage develops there that depends on the temperature at exactly this junction.
There is just one catch: they always measure only the temperature difference between the measuring junction and the connection point (terminals). To determine an absolute temperature, the temperature at the connection point must be known. In the past, this was solved with an ice bath, since it has a known temperature of 0 °C, which is where the term "cold junction compensation" comes from. In modern DAQ systems, a small temperature sensor (usually an RTD or thermistor) sitting directly at the connection point takes over this task. The software then calculates automatically: measured thermoelectric voltage + terminal temperature = actual temperature at the measuring point. In high-quality systems, CJC accuracy is ±0.5 °C.
There are many types of thermocouples, which differ in their metal pairings. Each pairing (type) has specific properties for temperature measurement. The most common types are:
Resistance thermometers are the standard for precise, stable measurements in labs, calibration, and industrial processes with high quality requirements.
Resistance thermometers, also called RTDs (resistance temperature detectors), use the property of metals that their electrical resistance changes with temperature. Platinum shows a highly reproducible, nearly linear characteristic.
A Pt100 has exactly 100 Ω at 0 °C and 138.5 Ω at 100 °C. A Pt1000 accordingly has 1,000 Ω at 0 °C. Above 0 °C, the behavior is approximately linear, which makes conversion into temperature values comparatively simple and accurate.
With Pt100 sensors, lead resistance distorts the result. With 10 meters of cable, this can quickly add up to an error of 2–3 °C, because the measuring current flows through the same two wires used for measurement. The lead resistance adds to the sensor resistance.
The solution, and in our view a must:
Use 3-wire or 4-wire connections. They separate the current paths so that only the actual sensor resistance is measured.
Thermistors suit applications that need high sensitivity, small size, and short response times, typically in electronics and devices.
A thermistor is a semiconductor resistor whose value changes strongly and reproducibly with temperature (described, e.g., by the Steinhart-Hart equation). In NTC thermistors, resistance falls as temperature rises; in PTC thermistors, it rises. This characteristic is used to measure temperature.
IR methods allow fast measurements without intervening in the process, or where the sensor cannot reach the measuring point. However, they capture only the surface temperature and depend heavily on the emissivity of the surface.
All bodies above absolute zero emit thermal radiation in the infrared range. An IR pyrometer or thermal imaging camera detects this radiation, filters out a specific wavelength band, and calculates the surface temperature from it. This is based on the Stefan-Boltzmann law and the emissivity of the surface.
Practical notes: The biggest problem with IR measurement: different surfaces radiate with different intensity. A pyrometer "sees" polished aluminum at 100 °C as much cooler than matte black sheet metal at the same temperature. A practical trick is to apply an "emissivity patch" with a known emissivity (often close to 1.0) or to mark a small spot with matte black paint (emissivity ≈ 0.95).
Compared with the sensors and methods described so far, fiber optic sensors are still relatively new, but for certain applications in EMC-critical, high-voltage, and explosion-hazard environments they are irreplaceable. They work without any electrical connection at the measuring point and are therefore intrinsically safe and immune to EMC. Two methods are currently used for temperature measurement with optical fibers:
A fiber Bragg grating (FBG) is inscribed into the glass fiber: a tiny grating that reflects light of a specific wavelength. This Bragg wavelength changes with temperature, because the refractive index and grating period depend on temperature. An interrogator (optical spectrometer) measures this shift and calculates the temperature. The temperature range depends on fiber and coating and is typically −40 to ~300 °C; special versions go higher.
Here, the intensity of the light passing through or reflected back from the fiber changes depending on external influences (e.g., temperature, absorption, bending). The light absorption or loss in the sensor is measured, and the temperature is derived from it.
The temperature is captured at the end of the fiber, similar to thermocouples or RTDs. This makes the method particularly suitable for R&D and testing when measurements are needed in bores or at precisely defined points. Because the fibers can be made extremely thin, they can be integrated into tight gaps, for example directly in the windings of electric motors. Conventional sensors often reach their limits there, if only because of their size and the insulation required.
Note: Besides the two methods described, there is also distributed fiber optic temperature sensing (DTS) via Raman/Brillouin scattering, which measures temperature continuously along a fiber, for example over long distances (e.g., pipeline monitoring). For test setups and test stands, FBG sensors are often the more practical solution because they measure temperature at a defined point (the grating).
For dynamic processes, thermocouples and thermistors often have the edge. If a wide temperature range is needed, thermocouples are a good choice. For precision and long-term use, RTDs offer the best basis. IR scores when contact is impossible. Fiber optics are recommended for EMC/HV environments, explosion protection, or when many measuring points along a route are needed. The following table gives quick orientation along key selection criteria and typical applications.
| Criterion | Thermocouple | RTD (Pt100/Pt1000) | Thermistor (NTC) | IR pyrometer | Fiber optic (FBG) | Fiber optic (intensity-based) |
|---|---|---|---|---|---|---|
| Temperature range | very wide (−200…>1,000 °C) | −200…~600 °C | typ. −40…150 °C | −50…>2,000 °C (material-dependent) | typ. −40…~300 °C (coating-dependent), special versions higher | typ. −40…~300 °C (depends on sensor design) |
| Accuracy | medium | high | medium–high (narrow range) | medium (ε-dependent) | high (with calibration/packaging) | solid, but sensitive to external influences |
| Response time | fast | medium | very fast | very fast | very fast (low mass) | very fast (end-point measurement) |
| Robustness | high | high | medium | high (optical) | high (electrical) medium (mechanical) |
high (electrical) medium (mechanical) |
| Cost | low | medium | low | higher | high (interrogator) | medium–high (simpler interrogator) |
| Typical applications | exhaust, furnace, engine parts | lab, calibration, process | smartphone, electronics | moving/hazardous parts such as brake discs | ATEX, HV batteries, transformers | single points in bores, motor windings, hazardous areas |
The best sensor is useless if there are errors in the measurement chain. Only the right measurement chain, with suitable signal conditioning, digitization, synchronization, and processing, turns sensor signals into reliable temperature data. One important principle is differential measurement, that is, between the + and − lines. This reliably suppresses interference that occurs on both lines at the same time (e.g., 50/60 Hz mains hum).
To use thermocouples reliably, the entire measurement chain from the sensor via the extension cable to the amplifier must be properly planned.
Cold junction compensation (CJC): Modern DAQ systems and instruments already have CJC built in. You just select the right thermocouple type in the software and enable CJC. Problems usually arise only with simple handheld meters, homemade electronics, or when the CJC function has been disabled by mistake.
With RTDs such as the Pt100, correct wiring and linearization above all determine the precision of the measurement chain.
Don't underestimate linearization:
RTDs are not linear! The Callendar-Van Dusen equation describes the real behavior. Many systems use simplified linear approximations, which leads to errors of several degrees over larger temperature ranges.
With non-contact methods, the optical chain from the measuring spot to evaluation is decisive for correct results. They are particularly suited to moving, hot, or hazardous components.
On a rotating shaft, the temperature is captured with a short-wavelength IR pyrometer. A matte marking prevents reflection errors.
FBG sensors work without an electrical connection and are suited to high voltage, strong EMC, and hazardous areas. Fiber optic sensors also need a matched measurement chain, from the sensor grating via the fiber to the interrogator.
FBG sensors respond to both temperature and mechanical strain. For pure temperature measurement, you need strain-compensated FBGs or mechanically decoupled mounting.
Application example: In an HV battery module, several FBGs are routed on one fiber along the cell rows. The interrogator captures all measuring points synchronously and galvanically isolated from the high-voltage system.
Regardless of sensor type, acquisition must deliver the bandwidth, isolation, and time base the application requires.
Helpful guidelines:
Important note: 24-bit resolution on the data sheet does not automatically mean high measurement quality. Optimize the analog foundation (sensor, amplifier, cabling) first, then make use of the digital capabilities.
There are recurring error sources that can be reliably controlled with a few principles.
Causes:
Solution:
Causes:
Solution:
Causes:
Solution:
Quick estimates support design and serve as plausibility checks during testing.
Rule of thumb: The sensor delivers approx. 40 µV (0.00004 V) per kelvin.
Example: A temperature difference of 100 °C corresponds to a 4 mV (0.004 V) signal.
Practical tip: The amplifier should amplify at least 10 times, giving you 0.4 mV/K, which is easy to measure.
Rule of thumb: approx. 0.4 Ω per kelvin
Example: At 100 °C, a Pt100 has about 138 Ω instead of 100 Ω at 0 °C
Practical tip: 1 mA measuring current → 0.4 mV/K signal change
Thermocouple (0.2 mm) in air: τ ≈ 0.1–0.5 s
Pt100 with protection tube in liquid: τ ≈ 1–10 s
Infrared pyrometer: τ < 0.1 s
τ (tau) = time until 63 % of a step change in temperature is captured.
The checklist helps you capture requirements in a structured way and narrow down options systematically.
Task: monitor oil temperature, coolant, exhaust gas
Solution:
Task: find hot spots on circuit boards
Solution:
Task: reactor temperature in a hazardous area
Solution:
Task:
Monitor cell temperature, cooling circuit, and thermal runaway management
Solution:
Rough guide (as of 2025):
Lead times:
There is no "best" temperature sensor, only the best one for your specific application:
The most important tip:
The measurement chain is decisive! The most expensive sensor is useless if cold junction compensation, amplification, or calibration are wrong.
For beginners: start with standard thermocouples or Pt100; they solve 90 % of all tasks. Special sensors such as FBG are for very specific problems.
YOUR NEXT STEP
We look forward to learning about your test and measurement tasks and developing solutions together that take your projects to the next level.
This website may use cookies for statistics and marketing purposes.
The following cookies may be set depending on your settings:
user_privacy_settings
| Domain name: | tmx-solutions.com |
| Expiry: | 30 days |
| Storage: | Local storage |
| Description: | Stores the privacy settings from the "Privacy Manager" consent tool. |
user_privacy_settings_expires
| Domain name: | tmx-solutions.com |
| Expiry: | 30 days |
| Storage: | Local storage |
| Description: | Stores the expiry of the privacy level settings from the "Privacy Manager" consent tool. |
ce_popup_isClosed
| Domain name: | tmx-solutions.com |
| Expiry: | 30 days |
| Storage: | Local storage |
| Description: | Stores that the user closed the popup (content element "Popup"). |
Eclipse.outdated-browser: "confirmed"
| Domain name: | tmx-solutions.com |
| Expiry: | 30 days |
| Storage: | Local storage |
| Description: | Stores the state of the "Outdated browser" notice bar. |